Epic Code LAB1230425 Complement C3, Serum
Additional Codes
Mayo Code: C3
Epic Code: LAB 1230425
Performing Laboratory
Mayo Clinic Laboratories in RochesterUseful For
Assessing disease activity in systemic lupus erythematosus
Investigating an undetectable total complement (CH50) level
Specimen Type
SerumSpecimen Required
Collection Container/Tube:
Preferred: Serum gel
Acceptable: Red top
Submission Container/Tube: Plastic vial
Specimen Volume: 1 mL
Collection Instructions: Centrifuge and aliquot serum into a plastic vial.
COLLECTION NOTE: Volumes listed are in serum or plasma, draw approximately 2 1/2 times the requested volume in whole blood.
Specimen Minimum Volume
0.5 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum | Refrigerated (preferred) | 7 days | |
Frozen | 28 days | ||
Ambient | 72 hours |
Reject Due To
Gross hemolysis | OK |
Gross lipemia | Reject |
Gross icterus | OK |
Day(s) Performed
Monday through Friday
Reference Values
75-175 mg/dL
Clinical Information
The complement system is an integral part of the body's immune defenses. It can be activated via immune complexes (classic pathway) or by bacterial polysaccharides (alternative pathway).The primary complement pathway consists of recognition (Clq, Clr, Cls), activation (C4, C2, C3), and attack (C5, C6, C7, C8, C9) mechanisms with respect to their role in antibody-mediated cytolysis.
C3 activation involves cleavage by C3 convertase into C3a and C3b. When immune complexes are not involved, the alternate method of complement activation initiates the reactant sequence at C3, bypassing C1, C4, and C2.
Severe recurrent bacterial infections occur in patients with homozygous C3 deficiency and in those patients with low levels of C3 secondary to the absence of C3b activator.
Decreased C3 may be associated with acute glomerulonephritis, membranoproliferative glomerulonephritis, immune complex disease, active systemic lupus erythematosus, septic shock, and end-stage liver disease.
Cautions
The results are dependent on appropriate specimen transport and storage.
Interpretation
A decrease in C3 levels to the abnormal range is consistent with disease activation in systemic lupus erythematosus.
Reporting Name
Complement C3, SMethod Name
Nephelometry
Method Description
C3 is measured by immunonephelometry. Antiserum to C3 is mixed with patient serum, the light scatter resulting from the antibody interaction with C3 is measured, and the signal is compared to standard concentrations of C3.(Instruction manual: Siemens Nephelometer II Operations. Siemens, Inc; Version 2.4, 07/2019; Addendum to the Instruction Manual 2.3, 08/2017)
CPT Code Information
86160
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
C3 | Complement C3, S | 4485-9 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
C3 | Complement C3, S | 4485-9 |
Report Available
Same day/1 to 2 daysTest Classification
This test has been cleared, approved, or is exempt by the US Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.Clinical Reference
1. Willrich MAV, Braun KMP, Moyer AM, Jeffrey DH, Frazer-Abel A: Complement testing in the clinical laboratory. Crit Rev Clin Lab Sci. 2021 Nov;58(7):447-478. doi: 10.1080/10408363.2021.19072972
2. Wong EKS, Kavanagh D: Diseases of complement dysregulation-an overview. Semin Immunopathol. 2018 Jan;40(1):49-64. doi: 10.1007/s00281-017-0663-8
3. Prohaszka Z, Kirschfink M, Frazer-Abel A: Complement analysis in the era of targeted therapeutics. Mol Immunol. 2018 Oct;102:84-88. doi: 10.1016/j.molimm.2018.06.001
4. Brodszki N, Frazer-Abel A, Grumach AS, et al: European Society for Immunodeficiencies (ESID) and European Reference Network on Rare Primary Immunodeficiency, Autoinflammatory and Autoimmune Diseases (ERN RITA) Complement Guideline: Deficiencies, Diagnosis, and Management. J Clin Immunol. 2020 May;40(4):576-591. doi: 10.1007/s10875-020-00754-1
Forms