Epic Code LAB1231383 Total Complement (CH50)
Additional Codes
Interface Order Alias
1231383
Clinical Information
Total Complement (CH50) is a functional assay that measures the activity of complement components C1 to C9 (i.e., the classical and terminal pathways). It does not measure the functional activity of the alternative pathway. If the Total Complement (CH50) measurement is zero or very low it is advisable to repeat the test to confirm the result. If the result is confirmed, the clinical presentation should be considered prior to further testing. If the patient exhibits recurrent neisserial infection, immunochemical determination of C5, C6, C7 and C8 is advised (or a functional assay for C5-C8). If the clinical problem is angioedema, initial studies should include C3, C4, C1 inhibitor (and possibly C1q and C2). Acquired disorders such as immune-mediated glomerulonephritis and other autoimmune disorders should direct testing towards C3 and C4.
Collection Instructions
Container(s): Red top (Serum gel/SST are not acceptable)
Preferred Volume to Collect: 6 mL
Minimum Volume to Collect: 2 mL
Submission Container/Tube: Plastic vial
Collection Instructions:
- Specimen should be processed immediately (within 1 hour) – See Processing Instructions
PROCESSING INSTRUCTIONS (Laboratory, Outpatient or Off-site collection)
Processed Specimen: Serum
Centrifuge/Spin: Yes
Aliquot: Yes
Processing Instructions:
- Allow blood to clot for 30 minutes in a vertical position and centrifuge promptly.
- Within an hour aliquot serum into a plastic transport tube and freeze as soon as possible.
- (+(1)S/PO/F)
Specimen Transport Temperature
Frozen
Rejection Criteria
- Specimens collected in any tube other than a plain Red-top tube.
- Specimens not collected and processed as indicated.
- Specimens not received frozen from off-site areas/offices.
- Severely hemolyzed, lipemic or icteric specimens.
Specimen Stability
Serum Specimens (Pour-Overs)
Frozen : 14 days
Test Frequency
Tuesday, Thursday.
Results available within 3 business days.
Reference Range
Greater than or equal to 41.7 U/mL.
Performing Department
REFERRAL SEND OUTS
Performing Department Laboratory Location
Beaumont Royal Oak Special Chemistry Laboratory
Methodology
Liposome Immunoassay (The Binding Site SPAPLUS Analyzer)
CPT
86162
CDM Code
3028616201
Epic Test ID
123010759
LOINC
4532-8
Reviewed Date
6/9/2023