Epic Code LAB1231960 Echovirus Antibodies
Test Name Alias
LAB1231191 | Quest: 35139 | ECHOVIRUS AB
Interface Order Alias
1231960
Clinical Information
Echovirus Antibody Panel - The current echovirus panel includes serotypes 4, 7, 9, 11, and 30. The echovirus serotypes used in this panel will vary depending upon current epidemiological data available from inhouse isolations CDC reports. Generally, the serotypes used are those most commonly associated with viral meningitis. Although there is cross-reactivity among the enteroviruses, most healthy people do not have detectable CF titers (>1:8). Therefore, detectable titers, especially those >1:32, should be considered diagnostic. Confirmed serodiagnosis is made by demonstration of a four-fold change in titers between acute and convalescent sera.
Specimen Requirements
Specimen Collection Criteria
Collect (preferred specimen): One Gold-top SST tube.
Also acceptable: One plain Red-top tube
Physician Office/Draw Specimen Preparation
Let specimen clot 30-60 minutes then centrifuge to separate serum from cells. Transfer serum to plastic transport tube and maintain at room temperature (20-26°C or 68-78.8°F) prior to transport.
In-Lab Processing
Let specimen clot 30-60 minutes then centrifuge to separate serum from cells. Transfer serum to plastic transport tube and maintain at room temperature (20-26°C or 68-78.8°F) prior to transport.
Specimen Transport Temperature
Transport: 2.0 mL serum , room temperature (20-26°C or 68-78.8°F)
Rejection Criteria
- Specimens not collected and processed as indicated.
Specimen Stability
Specimen Stability for Testing:
Room Temperature (20-26°C or 68-78.8°F): 7 days
Refrigerated (2-8°C or 36-46°F): 14 days
Frozen (-20°C/-4°F or below): 30 days
Specimen Storage in Department Prior to Disposal:
Specimen retention time is determined by the policy of the reference laboratory. Contact the Send out Laboratory with any questions.
Test Frequency
Performed: Monday – Friday.
Results available in 2-5 days.
Reference Range
|
Echovirus 4 Ab |
<1:8 titer |
|
Echovirus 7 Ab |
<1:8 titer |
|
Echovirus 9 Ab |
<1:8 titer |
|
Echovirus 11 Ab |
<1:8 titer |
|
Echovirus 30 Ab |
<1:8 titer |
Performing Department
Send Outs
Performing Department Laboratory Location
Sent to Quest Diagnostics- Wood Dale, IL
Methodology
Complement Fixation (CF
CPT
86658 x 5.
LOINC
Unavailable
Reviewed Date
11/2/2025
Orderable Regions
Send Outs