Epic Code LAB1232036 Phospholipase A2 Receptor (PLA2R) Antibody Panel
Test Name Alias
LAB1231561, LAB3489, LAB1231562, LAB1231771 | Quest: 93875 | Primary Membranous Nephropathy,Membranous Nephropathy,Glomerulonephritis,Anti-PLA2R,Idiopathic Membranous Nephropathy,PLA2R Antibodies
Interface Order Alias
1232036
Clinical Information
Phospholipase A2 Receptor (PLA2R) Antibody Panel - Phospholipase A2 Receptor (PLA2R) Antibody (Ab) testing provides physicians with a non-invasive tool for diagnosing primary membranous nephritis (PMN), an autoimmune disease most commonly mediated by autoantibodies to the podocyte antigen, PLA2R. PLA2R Ab is a highly specific biomarker for PMN and is detected in about 70-80% of patients. Moreover, PLA2R Ab levels frequently rise and fall with the clinical course of PMN and PLA2R Ab levels may have clinical utility in monitoring PMN patients for therapeutic response, remission, and disease relapse.
Specimen Requirements
Specimen Collection Criteria
Collect (preferred specimen): One Gold-top SST tube.
Also acceptable: One plain Red-top tube.
Physician Office/Draw Specimen Preparation
Let specimen clot 30-60 minutes then centrifuge to separate serum from cells. Transfer serum to a plastic transport tube and keep at room temperature (20-26°C or 68-78.8°F) prior to transport.
n-Lab Processing
Let specimen clot 30-60 minutes then centrifuge to separate serum from cells. Transfer serum to a plastic transport tube and keep at room temperature (20-26°C or 68-78.8°F) prior to transport.
Specimen Transport Temperature
1.5 mL serum, room temperature (20-26°C or 68-78.8°F) prior to transport. (Minimum: 1.0 mL)
Rejection Criteria
- Specimens not collected and processed as indicated.
Specimen Stability
Room Temperature (20-26°C or 68-78.8°F): 7 days
Refrigerated (2-8°C or 36-46°F): 14 days
Frozen (-20°C/-4°F or below): 21 days
Specimen Storage in Department Prior to Disposal:
Specimen retention time is determined by the policy of the reference laboratory. Contact the Send Outs Laboratory with any questions.
Test Frequency
Unavailable
Reference Range
Phospholipase A2 Receptor (PLA2R) Ab, ELISA
| RU/mL | |
| <14 | Negative |
| 14-19 | Borderline |
| >19 | Positive |
| Phospholipase A2 Receptor (PLA2R) Ab, IFA | Negative |
Performing Department
Send Outs
Performing Department Laboratory Location
Quest Diagnostics Nichols Institute-San Juan Capistrano, CA
Methodology
Enzyme Linked Immunosorbent Assay (ELISA) • Immunofluorescence Assay (IFA)
CPT
83520, 86255
LOINC
| PHOSPHOLIPASE A2 RECEPTOR (PLA2R) AB, ELISA | 73737-9 |
| PHOSPHOLIPASE A2 RECEPTOR (PLA2R) AB, IFA | 82991-1 |
Reviewed Date
11/17/2025