Epic Code LAB3227 Drug Screen Toxicology, Immunoassay, Urine
Test Name Alias
UDS | DrugScrAbuseTox | Drug of Abuse Screen w/alcohol | DOA | 4008
Interface Order Alias
11576
Collection Instructions
Specimen Collection: Urine, Random
Container(s): Preferred: Sterile Urine Cup or Pale Yellow Top Urine Tube-no preservative
Uscreen and iSCREEN POC (Point of Care) urine drug testing cups are acceptable but not preferred
Preferred Volume to Collect: 10.0 mL
Minimum Volume to Collect: 1.0 mL
Collection Instructions:
- Include Chain of Custody form when appropriate
- Keep container upright at all times and check that lid is fastened evenly and securely.
- Do not send container home with patient, must be collected on site.
- Do not open container after collection.
Processing Instructions (Laboratory, Outpatient or Off-site collection)
Transport Temperature: Refrigerate
Specimen Stability
Ambient: 72 hours
Refrigerate: 14 days
Frozen: 14 days
Laboratory Retention: 10 days
Reflex Information
Mandatory Reflex Information:
- Positive amphetamine, cannabinoids, methadone, opiates, oxycodone or cocaine for Inpatient Obstetric patients (mothers and their babies), LC/MS Confirmation will be performed.
Test Frequency
Available 24 hours, usual TAT 1 day.
STAT TAT 1 hour.
Reference Range
Screen Cutoff Limits:
Amphetamines 500 ng/mL
Barbiturates 200 ng/mL
Benzodiazepines 300 ng/mL
Cannabinoids 50 ng/mL
Cocaine/Metab. 150 ng/mL
Ethanol 20 mg/dL
Methadone 300 ng/mL
Opiates 300 ng/mL
Oxycodone 100 ng/mL
SCREENING PROCEDURE ONLY. Positives are not confirmed except where established by Laboratory policy. For other specimens, confirmation testing must be ordered by the physician. The Laboratory recommends that no legal action be taken on a positive screen result without confirmatory testing being performed.
Screening procedures generally target only one or a few specific drugs of a given class. Other drugs of the class may or may not cross react. Comprehensive Drug Screen may detect drugs not detected by this screening procedure.
pH and SG testing are performed to evaluate specimen integrity. Results outside the indicated ranges may or may not indicate an attempt by the patient to adulterate or otherwise compromise the specimen. Repeat testing on a new collection may be indicated if there is a question as to specimen integrity.
All urine drug screen specimens are stored for 14 days. Contact the Toxicology laboratory at 267-2780 for questions about confirmation testing, cross reactivity, Comprehensive Drug Screen, or specimen integrity testing.
Creatinine UR
Samples with creatinine <20 mg/dL may be too dilute for accurate drug detection.
pH Screen UR
4.00 – 8.50
SG Screen UR
1.003 – 1.030
Performing Department
Toxicology
Performing Department Laboratory Location
Corewell Health Reference Laboratory, Grand Rapids, MI
Methodology
Kinetic Interaction of Micro particles in a Solution (KIMS) / Enzyme Immunoassay (EIA)
CPT
80307
CDM Code
3018030716
Epic Test ID
1230100636
Mayo Access Code
SHO11576
Reviewed Date
2/15/2023