Epic Code LAB493 Protein S Antigen Free and Total
Test Name Alias
Protein S Ag | 4175 | Protein S Antigen | SHO4175 | 1230101091
Interface Order Alias
10461
Specimen Requirements
Specimen Type: Platelet Poor Plasma
Container/Tube: Blue top (3.2% Sodium Citrate)
Volume: 2.7 mL Whole Blood (Tube must be full) 2 tubes required
Minimum volume: 2 - 1.0 mL aliquots
Collection Instructions: Centrifuge and remove plasma within 1 hour of collection. Freeze aliquots in plastic transport tubes immediately
Specimen cannot be hemolyzed or clotted.
To obtain a coagulation plasma sample, centrifuge the capped specimen tube at a speed and time required to consistently produce platelet-poor plasma (platelet count < 10,000/uL). The centrifugal speed and duration to accomplish this must be established in the individual laboratory. Our laboratories use centrifuges at 3,000 to 4,000 rpm for ten minutes or higher speed and shorter duration centrifuges. When removing plasma to send for coagulation testing, it is important NOT to go all the way down to the cells. Leave a minimum 1/4 inch of plasma sitting on cells. This layer contains the platelets which were separated from the plasma during spinning. The platelets interfere with coagulation testing so we leave this layer undisturbed.
Specimen Stability
Ambient: 1 hour
Frozen: 2 months
Test Frequency
Sent to reference laboratory. Usual TAT 7-10 days.
Performing Department
Send Outs
Performing Department Laboratory Location
Send Out to Versiti Wisconsin Ref. Lab
CPT
85306, 85305
CDM Code
80101922, 80101926
Epic Test ID
1230101091
LOINC
Protein S Antigen Total: 5892-5
Protein S Antigen Free: 4677-1
Mayo Access Code
SHO4175
Reviewed Date
6/9/2023
Orderable Regions
Send Outs