Sign in →

Epic Code LAB7140 Ethylene Glycol

Test Name Alias

Ethylene Glycol | 4073

Interface Order Alias

10372

Collection Instructions

Specimen Collection: Blood

 

Container(s): Gray Top (Potassium Oxalate and Sodium Fluoride) or Red top (no additive)

Preferred Volume to Collect: 3.0 mL

Minimum Volume to Collect: 1.0 mL

Neonate Volume to Collect: 1.0 mL

Capillary collect ok? Yes

Microtainer acceptable?  Yes

 

Collection Instructions:

  • After collection, gently invert tube 8-10 times.
  • Gray Top is preferred tube.
  • Specimen should be processed – See Processing Instructions.

Processing Instructions (Laboratory, Outpatient or Off-site collection)

Processed Specimen: Plasma or Serum

Centrifuge/Spin: Yes

Aliquot: Yes

 

Processing Instructions:

  • Gray top:
    • Centrifuge after collection
    • Pipette plasma into aliquot tubes
  • Red top:
    • When tubes are clotted (30-45 minutes), centrifuge specimens
    • Pipette serum into aliquot tube

Minimum plasma/serum volume: 0.5 mL

 

Transport Temperature: Ambient

Specimen Stability

Ambient: 72 hours

Refrigerate: 7 days

Frozen: 14 days

Laboratory Retention: 7 days

Test Frequency

Available 24 hours, usual TAT 1 day.

STAT TAT 2 hours.

Reference Range

Negative (detection limit 10mg/dL)

Critical Value: Laboratory Critical Values Table

Performing Department

Toxicology

Performing Department Laboratory Location

Corewell Health Reference Laboratory, Grand Rapids, MI

Methodology

Gas Chromatography-Flame Ionization Detector (GC-FID)

 

This test was developed and its performance characteristics determined by Spectrum Health Toxicology Laboratory. It has not been cleared or approved by the FDA. The laboratory is regulated under CLIA as qualified to perform high-complexity testing. This test is used for clinical purposes. It should not be regarded as investigational or for research.

 

 

CPT

82693

CDM Code

3018269301

Epic Test ID

1230100664

LOINC

Ethylene Glycol: 5646-5

Mayo Access Code

SHO122

Additional Information

Gold Top SST not acceptable

Reviewed Date

2/15/2023

Updated Date

10/13/2023 Reference Range

5/20/2024 Clinical Information removed