Epic Code LAB848 Factor VIII Inhibitor
Test Name Alias
Factor 8
Specimen Requirements
Specimen Type: Platelet Poor Plasma
Container/Tube: Blue top (3.2% Sodium Citrate)
Volume: 4.5 mL Whole Blood (Tube must be full)
Minimum volume: Two (2) separate 1.0mL platelet poor citrated plasma -frozen aliquots
Collection Instructions: Centrifuge and remove plasma within 1 hour of collection. Freeze aliquots in plastic transport tubes immediately
Special Instructions
Specimen cannot be hemolyzed or clotted.
To obtain a coagulation plasma sample, centrifuge the capped specimen tube at a speed and time required to consistently produce platelet-poor plasma (platelet count < 10,000/uL). The centrifugal speed and duration to accomplish this must be established in the individual laboratory. Our laboratories use centrifuges at 3,000 to 4,000 rpm for ten minutes or higher speed and shorter duration centrifuges. When removing plasma to send for coagulation testing, it is important NOT to go all the way down to the cells. Leave a minimum 1/4 inch of plasma sitting on cells. This layer contains the platelets which were separated from the plasma during spinning. The platelets interfere with coagulation testing so we leave this layer undisturbed.
Rejection Criteria
Hemolysis
Specimen Stability
Ambient: 1 hour
Frozen: 2 months
Test Frequency
Sent to reference laboratory. Usual TAT 7-10 days.
Performing Department
REFERRAL SEND OUTS
Performing Department Laboratory Location
Send Out to Versiti Wisconsin Ref. Lab
CPT
85335
Reviewed Date
6/9/2023