Epic Code LAB8870 Hepatitis C Virus RNA Quantitation by RT-PCR
Test Name Alias
Hepatitis C VirusRNA Quantitative | 4185
Interface Order Alias
10170
Quick Collect
+(4)S/PO/F
Ordering Instructions
If one or more of the following tests are to be ordered, only one EDTA plasma tube or one serum tube (minimum volume 2 mL) needs to be provided:
- Hepatitis C Virus RNA Quantitation by RT-PCR
- Hepatitis C Genotype with Amplification
In addition, positive HCV specimens will be stored by the laboratory for 1 month, so additional follow-up testing can be added to the original order by FAXing the completed “Additional/Add-on Laboratory Request” form to Spectrum Health Lab Central at 616-267-2751.
Collection Instructions
Specimen Collection: Blood
Container(s): 6 mL Pink Top (EDTA) OR 5 mL Gold top (Serum Separator-SST Gel)
Preferred Volume to Collect: Full tube (6 mL or 5 mL)
Collection Instructions:
- Gently invert EDTA tube(s) 8 - 10 times after collection.
- Specimen should be processed within 4 hours – See Processing Instructions.
Processing Instructions (Laboratory, Outpatient or Off-site collection)
- Processed Specimen: Plasma or Serum
- Centrifuge/Spin: Yes
- Aliquot: Yes
- Processes Minimum Volume: 2.0 mL Plasma or Serum
- Processing Instructions:
- Plasma is preferred specimen type
- With room temperature storage, centrifuge whole blood in EDTA or Serum Separator tube within 4 hours of collection.
- With refrigerated temperature storage, centrifuge whole blood in EDTA or Serum Separator tube within 72 hours of collection.
- Transfer plasma or serum to a screw-capped transport tube.
- Plasma or serum needs to be stored frozen in screw-capped transport tube after separation from cells
- Transport temperature: Prefer that plasma or serum be frozen for transport
Specimen Stability
Ambient: Stability of whole blood in EDTA or a Serum Separator tube is 4 hours
Refrigerate: 3 days
Frozen: 60 days
Laboratory Retention: 1 month
Reflex Information
- Specimens with Positive Hepatitis C Antibody results will be automatically reflexed to the quantitative Hepatitis C Virus RNA test
Test Frequency
Test performed twice weekly, usual TAT 3 - 5 days
Reference Range
HCV
Not Detected: <12 IU/mL
Linear range of positive results
Positive: 12 to 100,000,000 (1e8) IU/mL
Equivocal: Detected but below lower limit of quantitation (<12 IU/mL)
Performing Department
Molecular Diagnostics
Performing Department Laboratory Location
Corewell Health Advanced Technology Laboratory (ATL), Grand Rapids, MI
Methodology
Quantitative Reverse Transcriptase Real-Time Polymerase Chain Reaction (RT-PCR)
CPT
87522
CDM Code
3068752201
Epic Test ID
1230100767
LOINC
HCV RNA Result : 11259-9
HCV RNA Viral Load : 11011-4
Mayo Access Code
SHO4185
Reviewed Date
6/22/2023