Epic Code LAB3534 ACTH Level
Test Name Alias
Adrenocorticotropic Hormone | 6662
Interface Order Alias
11788
Quick Collect
PCT/ON ICE+(1)S/PO/F
Clinical Information
Useful for determining the cause of hypercortisolism and hypocortisolism
Ordering Instructions
Note: Routine ACTH does not need to be scheduled.
Grand Rapids: For scheduling ACTH Stimulation testing, please call the Spectrum Health Infusion Clinic at 616-486-6099 in Grand Rapids.
United Hospital: For scheduling ACTH Stimulation testing, contact the Cancer Center at 616-225-9330.
Collection Instructions
Specimen Collection: Blood
Container(s): Pre-Cooled, Plastic Lavender EDTA
Preferred Volume to Collect: 3.0 mL
Minimum Volume to Collect: 1.0 mL
Neonate Volume to Collect: 1.0 mL
Capillary collect ok? No
Collection Instructions:
- PCT: Only use pre-chilled tubes
- Morning collection is optimal (6 am - 10 am)
- ON ICE: After collection immediately put specimen on ice
- Specimen should be processed immediately.
Processing Instructions (Laboratory, Outpatient or Off-site Collection):
Processed Specimen: Plasma (0.5 mL)
Centrifuge/Spin: Yes*
Aliquot: Yes
Processing Instructions:
- +(1)S/PO/F: Within one hour of collection, spin at 3500 for 5 minutes* or 3300 for 10 minutes in a prechilled red holder
- *Use refrigerated centrifuge to separate the plasma.
- Aliquot and freeze immediately.
Transport Temperature: Frozen
Specimen Stability
Refrigerated (separated plasma): 3 hours
Frozen: 4 weeks
Laboratory Retention: 72 hours
Test Frequency
Available Monday, Wednesday & Friday, TAT 24 hours
Reference Range
10 - 60 pg/mL (a.m. draws) No established reference values for p.m. draws.
Critical Values Table (for ages 0 - 18 years)
Performing Department
Chemistry
Performing Department Laboratory Location
Corewell Health Reference Laboratory, Grand Rapids, MI
Methodology
Method is Roche cobas Electrochemiluminescense Immunoassay.
Patient results determined by assays using different manufacturers for methods may not be comparable.
CPT
82024
CDM Code
3018202401
Epic Test ID
1230100301
LOINC
Not available
Reviewed Date
3/7/2024